About Us01
-
Business02
-
Product03
-
Licenses · Certifications04
-
Customer05
Description | Specification |
---|---|
Measuring Method |
Pyrocatechol Violet |
Measuring Parameter |
AI |
Measuring Range |
0~0.5/1.0/2.0/5.0 mg/L |
Detection Limit |
Lower than 0.003 mg/L (at 580 nm) |
Detector System |
580 nm Dual beam photometer |
Description | Specification |
---|---|
Measuring Method |
Formaldoxime |
Measuring Parameter |
Mn |
Measuring Range |
0~1.0/5.0/30.0 mg/L |
Detection Limit |
Lower than 0.004 mg/L (at 450 nm) |
Accuracy | Lower than ±3% |
Detector System |
450 nm Dual beam photometer |
Description | Specification |
---|---|
Measuring Method |
- TPTZ - Ferrozine |
Measuring Parameter |
Total Fe |
Measuring Range |
0~1.0/5.0/8.0/20 mg/L |
Detection Limit |
Lower than 0.0003 mg/L |
Accuracy | Lower than ±3% |
Detector System |
- 610 nm Dual beam photometer (TPTZ) - 560 nm Dual beam photometer (Ferrozine) |
Description | Specification |
---|---|
Measuring Method |
Zincon Method |
Measuring Parameter |
Copper |
Measuring Range |
0~1.0/2.0/5.0 mg/L |
Detection Limit |
Lower than 0.010 mg/L (at 620 nm) |
Detector System |
620 nm Dual beam photometer |
Description | Specification |
---|---|
Measuring Method |
Zincon Method |
Measuring Parameter |
Zine |
Measuring Range |
0~1.0/2.0/5.0 mg/L |
Detection Limit |
Lower than 0.010 mg/L (at 620 nm) |
Detector System |
620 nm Dual beam photometer |
Description | Specification |
---|---|
Measuring Method |
Chrono Potentiometry |
Measuring Parameter |
Pb, Cu |
Measuring Range |
0~1,000 ㎍/L … 0~2.0 mg/L |
Detector System |
Potentiostat |
Analytical Method |
Linear, Difference Pulse, Square Wave, Cyclic Voltammetry |
Description | Specification |
---|---|
Measuring Method |
Zincon Method |
Measuring Parameter |
Copper / Zinc |
Measuring Range |
0~1.0/2.0/5.0 mg/L |
Detection Limit |
Lower than 0.010 mg/L (at 620 nm) |
Detector System |
620 nm Dual beam photometer |